| Description: |
With a simple four button interface for easy instrument set-up.This versatile detector can meet the demands of all HPLC applications.
The Model 300S provides both heating and sub-ambient cooling of the Thermo-Split chamber for superior ELS detection at a price that meets tight budgets.
Features:
★ Superior HPLC Detection to Replace or Complement your Current Detection★ Sensitive Detection for a Wide Variety of Chemical Classes and Structures
★ Patent Pending Thermo-Split™ Technology for Precise Vapor Phase Control
★ Easy to Use
★ Fast Start-up
★ No Consumable Parts
★ Low Cost of Operation
Specification
|
Model 300s |
|
Dimensions |
9.8” W x 18” D x 11.5” H |
|
Weigh |
25 lbs |
|
Display |
2 Line x 20 Character per line VFD |
|
User Interface |
Four multi-function button |
|
Evaporative Zone Temperature |
Ambient to 120°C |
|
Thermo-Split™ Chamber Temperature |
10°C to 70°C |
|
Liquid Flow Rate |
Up to 5mL/min |
|
Gas Requirements |
60 - 80 psi Nitrogen or other inert gas |
|
Gas Consumption |
~ 2.5 SLPM |
|
Operating Conditions |
Intended for indoor use only, 60°F to 85°F and <90% R.H. non condensing |
|
Electrical Requirements |
Nominal 120 VAC, 50/60 Hz or Nominal 240 VAC, 50/60 Hz; 600 watts |
|
Wetted Materials |
Stainless steel, glass, anodized aluminum, Teflon™ |
|
Light Source |
650 nm Laser Diode, <5mW |
|
Detector |
Hermetically sealed photo-diode/operational amplifier |
|
Output Signal |
10mV, 5V, choosed by the Users |
|
Interface |
RS232, Contact Closure |
|